RadiCS application class | Body region/methods | Maximum display luminance cd/m2 | Maximum luminance ratio | Matrix of the image rendering device | Luminance characteristic curve | Homogeneity of the image rendering device | Homogeneity between multiple devices |
I. | Mammography | >=250 | >=250:1 | >=2048 x >=2048 | GSDF +/- 10% | <25 % | <10 % |
II. | Mammographic stereotaxis | >=200 | >=100:1 | >=1024 x >=1024 | GSDF +/- 15% | <25 % | <20 % |
III. | Projection radiography (thorax, skeleton, abdomen) | >=250 | >=250:1 | >=1600 x >=1200 | GSDF +/- 10% | <25 % | <20 % |
IV. | Fluoroscopy, all applications | >= 150 | >= 100:1 | >=1024 x >=1024 | GSDF +/- 15% | <25 % | <20 % |
V. | Computer tomography | >= 150 | >= 100:1 | >=1024 x >=1024 | GSDF +/- 15% | <25 % | <20 % |
VI. | For RK 5: Dental digital volume tomography intraoral X-ray diagnostics with dental X-ray tube heads, panoramic tomographs, teleradiograph exposures of the cranium, dental cranial overview exposures, hand radiographs to detect skeletal growth | >= 200 | >= 100:1 | >=1024 x >=768 | - | <30 % | <30 % |
VII. | For RK 6: Intraoral X-ray diagnostics with dental X-ray tube heads, panoramic tomographs, teleradiograph exposures of the cranium, dental cranial overview exposures, hand radiographs to detect skeletal growth | >= 300 | >= 100:1 | >=1024 x >=768 | - | <30 % | <30 % |
VIII. | Clinical review | - | - | - | GSDF | - | - |